Spesifikasi :
Features
• Visual and alarm sounds for already weakened batteries, infusion process has been completed, valves open, incorrect settings etc;
• HD LCD display, Easy way of using, dynamically displaying working status;
• Compatible with brands of infusion sets after correct calibration
• Preset volume solution that greatly reduces the workload of nurses
• Mode selection: ml / hr and drops / min, and both modes can be freely changed;
• Three levels of occlusion: high, medium and low;
• Cleaning function
• KVO (Keep Vena Open) vein remains open, automatically when the infusion is complete
• Power source: 100 ~ 240Vac 50 / 6OHz: Internal battery
• Automatically record last infusion settings (infusion pump with drug library can store 2000pcs infusion note paint)
Infusion Specifications
- Infusion rate 1ml / h ~ 1,200 miles / hour
- Stream flow accuracy: ± 5% (after calibration)
- Mechanical precision: ± 2%
- Cleaning rate: 100ml / h-1,200ml / hour
- KVO (Keep Vena Open) Level: 1ml / h ~ 5 ml / hour (1ml / hour per step)
- Infusion volume: 1ml ~ 9999 ml
- Total Infusion Volume: 0.1 ml – 9999.9 ml
- Occlusion: height: 800mmHg 1200mmHg (106.7kPa: 26.7kPa)
- medium: 500mmHg i100mmHg (66.7kPal13.3kPa)
- low: 300mmHg 1100mmHg (40.7kPal13.3kPa):
- AlarmTube tube blocked; There are bubbles; closed door; Infusion is complete; Incorrect setting, low battery, AC power has pulled out; Clean; Data recovered; BOLUS data processing runs; Forgot to operate, etc.
- Power source: 100 ~ 240Vac, 50 / 60Hz; Internal Li battery, capacityz1, 600mAh, 11.1V voltage, 4 hours backup time at 25ml / hour rate (mid level designated ll’t GB 9706.27-2005) for new battery can for 12hours.
- Bubble Detector: Ultrasonic wave detector; Fuse detection sensitivity ≥25μL:
- Fuse F2AL / 250V, 2 pcs installed:
- 3OVA power consumption
- IP classification: lPX4
- Classification of equipment: class ll, internal power supply, type CF
- Condition: operating temperature: + 5 ° C ~ 400C;
• Relative humidity: 20% -90%
• atmospheric pressure: 86.0kPa ~ 106.0kPa - Transportation & storage conditions: temperature: -3O0C ~ + 550C
- Relative humidity ≤95%
- Dimensions: 140 mm (L) x 157 mm (W) x 220 mm (H)
- Net weight: 1.8 kg